Pt Coated probe 镀白金导电探针(AFM 探针)

Typical radius of uncoated tip : 8 nm
Resulting tip radius with the coating : < 30 nm
Full tip cone angle* : 40°
Total tip height : 12 - 18 µm
Probe material : n-type silicon
Tip coating : Platinum
Detector coating : Platinum

The Cr-Au coating is formed as a 30 nm Au film with a 20 nm Cr sublayer, which is deposited for better adhesion of Au. The coating is formed on both tipside and backside of the cantilever. The coating may cause cantilever bending within 3°.

APPLICATION

The gold coating is inert, which makes these probes applicable for many experiments in biology and chemistry. Gold coated probes are also used in certain electrical applications. 


1-CANTILEVER SERIES

Cantilever Resonance Frequency, kHz Force Constant, N/m
min typical max min typical max
14 series 110 160 220 1.8 5.0 13
15 series 265 325 410 20 40 80
17 series 10 13 17 0.06 0.18 0.40
18 series 60 75 90 1.2 2.8 5.5

 

3-CANTILEVERS SERIES

Cantilever Resonance Frequency, kHz Force Constant, N/m
min typical max min typical max
35 series Cantilever A 130 205 290 2.7 8.9 24
Cantilever B 185 300 430 4.8 16 44
Cantilever C 95 150 205 1.7 5.4 14
36 series Cantilever A 30 90 160 0.1 1.0 4.6
Cantilever B 45 130 240 0.2 2 9
Cantilever C 25 65 115 0.06 0.6 2.7
37 series Cantilever A 30 40 55 0.3 0.8 2
Cantilever B 15 20 30 0.1 0.3 0.6
Cantilever C 20 30 40 0.1 0.4 1

 

4-CANTILEVERS SERIES

Cantilever Resonance Frequency, kHz Force Constant, N/m
min typical max min typical max
11 series Cantilever A 12 15 18 0.1 0.2 0.4
Cantilever B 60 80 100 1.1 2.7 5.6
Cantilever C 115 155 200 3 7 16
Cantilever D 250 350 465 17 42 90